After many studies and field tests, it was found that the guide pins on the DSX-5000 and DSX-8000 copper modules are redundant. Tests with units with no guide pines have shown no variance in performance compared to tests with units with guide pins.
In a Service Alert issued in January of 2019, the worldwide service centers were advised that if they come across loose guide pins during repair or calibration procedures, the pins should be removed from the copper modules. It is uncommon for the pins to come loose. If they are still well connected, no action will be taken by the service center.
經(jīng)過多次研究和現(xiàn)場測試,發(fā)現(xiàn)dsx-5000和dsx2-8000銅模塊上的導銷是冗余的。與帶有導向銷的單元相比,沒有導向銷的單元的測試在性能上沒有差異。
福祿克維修中心在2019年1月發(fā)布告知,如果在維修或校準過程中遇到松動的導銷,應將導銷從銅模塊上拆下。別針松動是很少見的。如果仍然連接良好,服務中心將不采取任何行動。